Introduction
Capnograph adopts sidestream method to measure the end-tidal carbon dioxide, airway respiration rate and other parameters. It can detect pulmonary ventilation function and reflect the circulation and pulmonary blood flow as well as indirectly reflect the alveolar ventilation standard. Meanwhile, it can also detect the blood oxygen saturation and pulse.
The Capnograph is applicable for use in pre-hospital emergency transfer, emergency department, ICU, operating room, neonatology, respiratory department and other occasions. The application population is adult, pediatric and neonate patients.
Features
-
End-tidal CO2 measurement function
-
Respiration rate measurement function
-
SpO2measurement function
-
Pulserate measurement function
-
Over-limit alarm function for EtCO2、Respiration rate,SpO2and Pulse rate
-
Low battery alarm function
Performance
-
End-tidal CO2 measurement:
- Measurement range: 0 ~150mmHg
- Resolution: 1mmHg
- Accuracy:
0~40 mmHg:±2 mmHg
41~70 mmHg:±5% of readings
71~100 mmHg:±8% of readings
101~150 mmHg:±10% of readings
-
Respiration rate measurement:
- Measurement range: 2rpm~120rbpm
- Resolution: 1rpm
- Accuracy: ±1rpm
-
SpO2measurement:
-
Measurement range: 30~100 %
-
Resolution: 1%
-
Accuracy:
70~100 %:2 %
0 ~ 69 %:undefinition
-
Pulserate measurement:
-
Measurement range: 25~250 bpm
-
Resolution: 1bpm
-
Accuracy: ±2 bpmor ±2%,whichever is greater
-
Safety characteristic:
- Device type: internally powered equipment
- Degree of protection against electric shock: type BF applied part
- Degree of protection against ingress of liquid: IPX1
Accessories
-
User Manual
-
Blood oxygen probe
-
Sampling pipe
Physical Characteristic
Dimension: 159*72*31 mm
Weight: 270 g
Working environment
Temperature: 5℃~40℃
Relative humidity: 30%~ 75%, no condensation
Atmospheric pressure: 700hPa~1060hPa
Power supply: DC 3.7V (rechargeable lithium battries)
Input power: ≤1.5VA