1 Introduction
The monitor is applicable for clinical monitoring in operating room, postoperative observation room, ICU/CCU ward, emergency room, etc. of adult, pediatric and neonate. The patient's vital parameters (ECG (including ST-segment measurement and arrhythmia analysis), RESP, SpO2, PR, NIBP, TEMP, IBP and CO2, etc.) can be monitored. Monitored information can be displayed, reviewed, printed and stored..
2 Function
-
13.3" high-definition touch screen design, easy and convenient to operate, elegant and concise in appearance.
-
Independent physiological and technical alarm function, convenient for medical staff to quickly know patient’s condition.
-
Luminous “ON / OFF” button design, convenient to operate at night.
-
Modular design: the modules can be configured and combined flexibly and conveniently, expanded and upgraded at any time, meets the monitoring requirements of different departments on different parameters.
-
Support dynamic hot-plugging parameter module: no need to restart the device, plug-and-play, change the monitoring parameters at any time as clinical requirements.
-
Adopt high-performance processor, stable, reliable and fast to process information.
-
Patient information can be rapidly entered, patient type (adult / neonate / pediatric) can be quickly switched, meets the requirements of different departments.
-
Abundant analysis functions: 23 arrhythmia, full-lead ST-segment and pacing analysis; multiple calculation functions (such as calculation of drug concentration, titration table, hemodynamics, ventilation, oxygenation, renal function, etc.).
-
With the functions of waveform freezing, holographic waveform storage and review.
-
Automatic data storage in case of power failure; optional built-in memory card, achieves large-capacity data storage.
-
With HDMI expansion interface, the display screen can be connected according to clinical requirements.
-
Built-in large-capacity detachable lithium battery, ensures uninterrupted monitoring.
-
Fanless design, ensures no noise and low power consumption during working.
-
3-channel built-in recorder used to print real-time waveform is optional.
-
Connect to Central Monitoring System developed by our company by WiFi / Wired, convenient to transmit patient’s real-time data.
-
Standby mode, be applied to ICU, avoids affecting patient’s rest, one-button switching, reduces the workload of medical staff.
3 Performance
3-lead: I, II, III
5-lead: I, II, III, aVR, aVL, aVF, V
12-lead: I, II, III, aVR, aVL, aVF, V1,V2,V3,V4,V5,V6
-
Gain: 5mm/mV, 5.0 mm/mV, 10 mm/mV, 20 mm/mV, 40 mm/mV
-
Scan speed: 5 mm/s, 25 mm/s, 50 mm/s
-
HR
Measurement and alarm range:
Adult: 15 ~ 300 bpm
Pediatric/neonate: 15 ~ 350 bpm
Resolution: 1 bpm
Accuracy: ± 1 % or ± 1 bpm, whichever is greater
Measurement and alarm range: - 2.0 mV ~ + 2.0 mV
Accuracy: - 0.8 mV ~ + 0.8 mV: ± 0.04 mV or ± 10 %, whichever is greater
Other range: unspecified
-
Arrhythmia analysis: 23 kinds
-
Pacemaker: yes
-
RESP
-
Measurement mode: R-F(RA-LL) impedance
-
Respiration rate:
Measurement range:
Adult: 0 rpm ~ 120 rpm
Pediatric/neonate: 0 rpm ~ 150 rpm
Resolution: 1 rpm
Accuracy: 7 ~ 150 rpm, ± 2 rpm or ± 2 %, whichever is greater;
0 ~ 6 rpm, unspecified
-
Apnea alarm:10 ~ 120 s, not alarm
-
Scan speed: 25 mm/s, 12.5 mm/s, 25 mm/s, 50mm/s
-
NIBP
-
Measurement mode: Oscillometry
-
Working mode: Manual/AUTO/Continuous
-
Measurement interval in AUTO mode: 1min, 2 min, 5 min, 3 min, 5 min, 10 min, 15 min, 20 min, 30 min, 1 h, 1.5 h, 2 h, 3 h, 4 h, 8 h
-
Measurementperiod in Continuous mode: 5 minutes
-
Measurement and alarm range: 10~ 270 mmHg
-
Resolution: 1mmHg
-
Cuff pressure accuracy:± 3 mmHg
-
Measurement accuracy:
Maximal mean deviation: ± 5 mmHg
Maximal standard deviation: 8 mmHg
-
Over-pressure protection:
Adult mode: 297 ± 3 mmHg
Pediatric mode: 240 ± 3 mmHg
Neonatal mode: 147 ± 3 mmHg
-
CONTEC SpO2
-
Measurement range: 0% ~ 100 %
-
Resolution: 1%
-
Accuracy: 70% ~ 100 %, ± 2 %;
0 % ~ 69 %, unspecified
Measurement and alarm range: 25 ~ 250 bpm
Accuracy: ± 2 bpm or ± 2 %, whichever is greater
-
NellcorSpO2
-
Measurement range: 1% ~ 100 %
-
Measurement accuracy: 70% ~ 100 %: ± 2 % (adult / pediatric)
70 % ~ 100 %: ± 3 % (neonate)
0 % ~ 69 %: unspecified
Measurement range: 20 ~ 300 bpm;
Measurement accuracy: 20 ~ 250 bpm, ±3 bpm;
251 ~ 300 bpm: unspecified;
Upper limit of alarm: upper limit - 2 ~ 300 bpm;
Lower limit of alarm: 20 bpm ~ lower limit + 2 bpm
-
MasimoSpO2
-
Measurement range: 1% ~ 100 %
-
Measurement accuracy:
70 % ~ 100 %: ± 2 % (adult / pediatric, under conditions of no motion), ± 3 % (under conditions of motion)
70 % ~ 100 %: ± 3 % (neonate, under conditions of no motion)
1 % ~ 69 %: unspecified
Measurement range: 25 ~ 240 bpm
Measurement accuracy: ± 3 bpm (under conditions of no motion), ± 5 bpm (under conditions of motion)
Upper limit of alarm: upper limit - 2 ~ 240 bpm
Lower limit of alarm: 25bpm ~ lower limit + 2 bpm
-
TEMP
-
Channel: dual-channel
-
Measurement range: 0~ 50 ℃
-
Resolution: 1℃
-
Accuracy: ±1 ℃
-
CO2
-
Measurement mode: mainstream or sidestream
-
CO2measurement range: 0 ~ 150 mmHg
-
CO2resolution: 1 mmHg: 0 ~ 69 mmHg
0.25 mmHg: 70 ~ 150 mmHg
0 ~ 40 mmHg: ± 2 mmHg
41 ~ 70 mmHg: ± 5 %
71 ~ 100 mmHg: ± 8 %
101 ~ 150 mmHg: ± 10 %
Sidestream: 2 ~ 120 rpm
Mainstream: 2 ~ 150 rpm
-
AwRRaccuracy: ± 1 rpm
-
Apnea alarm:yes
-
IBP
-
Channel: 4-channel
-
Label: ART, PA,CVP, RAP, LAP, ICP, P1, P2
-
Measurement and alarm range: -50~ 300 mmHg
-
Resolution: 1 mmHg
-
Accuracy: ±2 % or 1 mmHg, whichever is greater
-
Power supply: 100~ 240 V ~ 50/60 Hz
-
Safety class: class I, defibrillation-proof type CF applied part
-
Waterproof degree: IPX2