(1)ECG
Lead type:
3-lead: I, II, III
5-lead: I, II, III, aVR, aVL, aVF, V
12-lead: I, II, III, aVR, aVL, aVF, V1,V2,V3,V4,V5,V6
Gain: 2.5 mm/mV, 5.0 mm/mV, 10 mm/mV, 20 mm/mV, 40 mm/mV
Scan speed: 6.25mm/s,12.5 mm/s, 25 mm/s, 50 mm/s
HR
Measurement and alarm range:
Adult: 15 ~ 300 bpm
Pediatric/neonate: 15 ~ 350 bpm
Resolution: 1 bpm
Accuracy: ± 1 % or ± 1 bpm, whichever is greater
ST-segment detection
Measurement and alarm range: - 2.0 mV ~ + 2.0 mV
Accuracy: - 0.8 mV ~ + 0.8 mV: ± 0.04 mV or ± 10 %, whichever is greater
Other range: unspecified
Arrhythmia analysis: 23 kinds
Pacemaker: yes
(2)RESP
Measurement mode: impedance
Respiration rate:
Measurement range: 0 rpm ~ 150 rpm
Resolution: 1 rpm
Accuracy: 7 ~ 150 rpm, ± 2 rpm or ± 2 %, whichever is greater;
0 ~ 6 rpm, unspecified
Apnea alarm: 10 ~ 60 s
Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50mm/s
(3)NIBP
Measurement mode: Oscillometry
Working mode: Manual/AUTO/Continuous/Order
Measurement interval in AUTO mode: 1 min, 2 min, 2.5 min, 3 min, 5 min, 10 min, 15 min, 20 min, 30 min, 1 h, 1.5 h, 2 h, 3 h, 4 h, 8 h
Measurement period in Continuous mode: 5 minutes
Static pressure Measurement range: 0mmHg ~ 300 mmHg(0.0kPa ~ 40kPa)
Resolution: 1 mmHg
Cuff pressure accuracy: ± 3 mmHg
Measurement accuracy:
Maximal mean deviation: ± 5 mmHg
Maximal standard deviation: 8 mmHg
Over-pressure protection:
Adult mode: 297 ± 3 mmHg
Pediatric mode: 260 ± 3 mmHg
Neonatal mode: 147 ± 3 mmHg
(4)SpO2
Measurement range: 0 % ~ 100 %
Resolution: 1 %
Accuracy: 70 % ~ 100 %, ± 2 %;
0 % ~ 69 %, unspecified
PR
Measurement and alarm range: 25 ~ 250 bpm
Accuracy: ± 2 bpm or ± 2 %, whichever is greater
(5)TEMP
Channel: dual-channel
Measurement range: 0 ~ 50 ℃
Resolution: 0.1 ℃
Accuracy: ± 0.1 ℃
(6)CO2
Measurement mode: mainstream or sidestream
CO2 measurement range: 0 ~ 150 mmHg
Accuracy:
0 ~ 40 mmHg: ± 2 mmHg
41 ~ 70 mmHg: ± 5 %
71 ~ 100 mmHg: ± 8 %
101 ~ 150 mmHg: ± 10 %
AwRR measurement range:
Sidestream: 2 ~ 120 rpm
Mainstream: 2 ~ 150 rpm
AwRR accuracy: ± 1 rpm
Apnea alarm: yes
(7)IBP
Channel: 2-channel(Expandable 4-channel)
Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measurement and alarm range: -50 ~ 300 mmHg
Resolution: 1 mmHg
Accuracy: ± 2 % or 1 mmHg, whichever is greater
(8)Power supply: 100V - 240 V ~ 50Hz/60 Hz
(9)Safety class: class I, defibrillation-proof type CF applied part
(10)Waterproof degree: IPX2