▲ECG
Lead mode: 3-lead: I, II, III
5-lead: I, II, III, aVR, aVL, aVF, V
Waveform: 3-lead, 1 channel
5-lead, 2-channel
Gain: 2.5 mm/mV, 5.0 mm/mV, 10 mm/mV, 20 mm/mV
Scan speed: 12.5 mm/s, 25 mm/s, 50 mm/s
HR
Measurement and alarm range:
Adult: 15 ~ 300 bpm
Pediatric/neonate: 15 ~ 350 bpm
Accuracy: ± 1 % or ± 1 bpm, whichever is greater
Resolution: 1 bpm
▲RESP
Method: R-F (RA-LL) impedance
Respiration rate:
Measurement and alarm range: 0 ~ 150 rpm
Resolution: 1 rpm
Accuracy:
7 rpm ~ 150 rpm, ± 2 rpm or ± 2 %, whichever is greater;
0 rpm ~ 6 rpm, unspecified
Apnea alarm: 10 ~ 40 s
Scan speed: 12.5 mm/s, 25 mm/s
▲NIBP
Measurement mode: Oscillometry
Working mode: Manual / AUTO / STAT
Measurement interval in AUTO mode: 1 / 2 /3 / 4 / 5/ 10 / 15 / 30 / 60 / 90 / 120 / 180 / 240 / 480 minutes
Measurement period in STAT mode: 5 minutes
Measurement and alarm range: 0 ~ 300 mmHg
Resolution: 1 mmHg
Cuff pressure accuracy: ± 3 mmHg
Measurement accuracy:
Maximal mean deviation: ± 5 mmHg
Maximal standard deviation: 8 mmHg
Over-pressure protection:
Adult: 297±3 mmHg
Pediatric: 260±3 mmHg
Neonate: 147±3 mmHg
▲SpO2
Measurement and alarm range: 0 ~ 100%
Resolution: 1 %
Accuracy: 70 % ~ 100 %: ± 2 %;
0 % ~ 69 %: unspecified
▲PR
Measurement and alarm range: 25 ~ 250 bpm
Accuracy: ± 2 bpm or ± 2 %, whichever is greater
▲TEMP
Measurement and alarm range: 0 ~ 50 ℃
Resolution: 0.1 ℃
Accuracy: ± 0.1 ℃
▲CO2(optional)
Measurement mode: mainstream or sidestream
Measurement range: 0 ~ 150 mmHg
Resolution:
0.1 mmHg: 0 ~ 69 mmHg
0.25 mmHg: 70 ~ 150 mmHg
Accuracy:
0 ~ 40 mmHg: ±2 mmHg
41 ~ 70 mmHg: ±5 %
71 ~ 100 mmHg: ±8 %
101 ~ 150 mmHg: ±10 %
AwRR measurement range:
Sidestream: 2 ~ 120 rpm
Mainstream: 2 ~ 150 rpm
Accuracy: ±1 rpm
Apnea alarm: yes
▲Power supply: 100 V-240 V ~ 50 Hz/60 Hz
▲Safety class: Class II
Physical characteristic
Dimension: 168 mm (L) × 56 mm (W) × 98 mm (H)
Weight: <1.0 Kg (excluding accessories)