1 Introduction
The monitor is used to monitor, display, check, store, alarm and transmit many physiological parameters, including ECG (3/5/12 lead, arrhythmia analysis, ST segment analysis and HR ), RESP, SpO2, PR, NIBP, TEMP and IBP, etc. It is applicable for use by adult, children and neonate.
The monitor can be used in medical institutions or during transport monitoring inside/outside the hospital environment, it is used by clinical professionals or under their guidance or professionals via training, it is not suitable for use in family.
2 Function
-
5" touch screen, easy and convenient to operate, automatic lock screen function prevents misoperation.
-
Portable design (weight <0 kg), flexible to transport.
-
Multi - interface selection, convenient to observe patient’s data trend.
-
Degree of protection provided by enclosure complies with IEC/EN 60529, waterproof degree: IP44
-
Fanless design, quiet, energy-saving and clean, which reduces the possibility of cross-infection.
-
Built-in large-capacity battery, meets the requirements of long-term operation.
3 Performance
3-lead: I, II, III
5-lead: I, II, III, aVR, aVL, aVF, V
3-lead: 1 channel
5-lead: 2-channel
-
Gain: 5mm/mV, 5.0 mm/mV, 10 mm/mV, 20 mm/mV
-
Scan speed: 5 mm/s, 25 mm/s, 50 mm/s
-
HR
Measurement and alarm range:
Adult: 15 ~ 300 bpm
Pediatric/neonate: 15 ~ 350 bpm
Resolution: 1 bpm
Accuracy: ± 1 % or ± 1 bpm, whichever is greater
Alarm error: ± 2 bpm
Measurement and alarm range: - 2.0 mV ~ + 2.0 mV
Accuracy: - 0.8 mV ~ + 0.8 mV: ± 0.04 mV or ± 10 %, whichever is greater
Other range: unspecified
-
Arrhythmia analysis: 13 kinds
-
Pacemaker: yes
-
RESP
-
Measurement mode: R-F(RA-LL) impedance
-
Respiration rate:
Measurement range: 0 ~ 150 rpm
Resolution: 1 rpm
Accuracy: 7 ~ 150 rpm, ± 2 rpm or ± 2 %, whichever is greater;
0 ~ 6 rpm, unspecified
-
Apnea alarm: 10~ 40 s
-
Scan speed: 12.5 mm/s, 25 mm/s
-
NIBP
-
Measurement mode: Oscillometry
-
Working mode: Manual/AUTO/Continuous
-
Measurement interval in AUTO mode: 1/ 2 / 3/ 4 / 5 / 10 / 15 / 30 / 60 / 90 / 120 / 240 /480 /960 minutes
-
Measurementperiod in Continuous mode: 5 minutes
-
Measurement and alarm range: 10~ 270 mmHg
-
Resolution: 1mmHg
-
Cuff pressure accuracy: ±3 mmHg
-
Measurement accuracy:
Maximal mean deviation: ± 5 mmHg
Maximal standard deviation: 8 mmHg
-
Over-pressure protection:
Adult mode: 297 ± 3 mmHg
Pediatric mode: 240 ± 3 mmHg
Neonatal mode: 147 ± 3 mmHg
-
SpO2
-
Measurement and alarm range: 0~ 100%
-
Resolution: 1%
-
Measurement accuracy: 70% ~ 100 %: ± 2 %;
0 % ~ 69 %: unspecified
-
PR
-
Measurement and alarm range: 25 ~ 250bpm
-
Accuracy: ±2 bpm or ± 2 %, whichever is greater
-
TEMP
-
Channel: dual-channel
-
Measurement and alarm range: 0 ~ 50 ℃
-
Resolution: 1℃
-
Accuracy: ±1 ℃
-
IBP
-
Channel: dual-channel
-
Label: ART, PA,CVP, RAP, LAP, ICP, P1, P2
-
Measurement and alarm range: -10 ~ 300 mmHg
-
Resolution: 1 mmHg
-
Accuracy: ±2 % or 1 mmHg, whichever is greater
-
Power supply: 100 ~ 240 V ~ 50 / 60 Hz (adapter output: DC 12 V)
-
Safety class: class I, defibrillation-proof type CF applied part
-
Waterproof degree: IP44
-
4 Accessories
Category
|
Accessory name
|
Main unit
|
TS1 main unit
|
ECG cable
|
5-lead, American standard, TPU
|
SpO2 probe (integrated)
|
Digital fingertip SpO2 probe for adult (CMS-N-SPO2 6P, 3M, yellow)
|
TEMP probe
|
R25=2.252K TEMP probe, skin surface type, CMS-N-TEMP 2P, PVC, L=3M
|
NIBP cuff
|
BP extension tube, L=3M
|
Adult cuff (25 ~ 35 cm)
|
-
5 Physical characteristic
- Dimension: 142 mm × 67 mm × 102 mm
-
Weight: < 1 kg
6 Optional configuration
IBP