Introduction
This device can monitor such parameters as ECG, RESP, SpO2, PR, NIBP and TEMP, etc. It integrates parameter measurement module, display and recorder in one device to form a compact and portable equipment.
Functions
Standard parameters: ECG, RESP, SpO2, PR, NIBP, single-channel TEMP
1)ECG Heart rate(HR)
ECG waveform
Arrhythmia and ST-segment analysis
2)RESP Respiration rate(RR)
Respiration waveform
3)SpO2 Oxygen saturation(SpO2)
Plethysmogram(PLETH) waveform
Pulse rate(PR)
Bar graph
4)NIBP Systolic pressure(SYS), Diastolic pressure(DIA), Mean pressure(MEAN)
5)TEMP Temperature
It has abundant functions, such as touch-screen operation, audible and visual alarm, trend data storage and output, NIBP measurement, alarm event marking and drug concentration calculation, etc.
Features
1)7'' TFT color LCD, multi-language interface(English, Chinese, Russian, Kazakhstan, Czech, Slovak).
2)Touch screen operation, which improves the working efficiency for medical staff.
3)Fanless design, quiet, energy-saving and clean, which reduces the possibility of cross-infection.
4)All-round monitor for adult, pediatric and neonate.
5)With standard interface, big character interface and view bed, convenient to observe.
6)Waveform, parameter color and location can be set optionally.
7)Display 7-lead ECG waveform on one screen.
8)Adopt digital SpO2 technology, anti-motion and anti-ambient light interference, and measurement can be performed under the circumstance of weak filling.
9)NIBP measurement mode: Manual/AUTO/Continuous, storage for 4800-group NIBP data.
10)Review for 71 alarm events of all parameters and 60 arrhythmia alarm events.
11)Drug concentration calculation and titration table functions.
12)Connect to Central Monitoring System by 3G, Wi-Fi or wired mode.
13)AC/DC, built-in rechargeable lithium battery achieve uninterrupted monitoring.
14)Anti-high frequency surgical unit, defibrillation-proof(special leads are necessary).
Performance
ECG
Lead mode: 3-lead or 5-lead
Lead selection: I, II, III, aVR, aVL, aVF, V
Waveform: single-channel
Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV
Scan speed: 12.5mm/s, 25 mm/s, 50 mm/s
HR:
Measurement and alarm range: 15~350 bpm
Accuracy: ±1 % or ±1 bpm, whichever is greater
Alarm accuracy: ± 2 bpm
Resolution: 1 bpm
ST-segment monitoring:
Measurement and alarm range: -2 mV ~ +2 mV
Accuracy: -0.8 mV~+0.8 mV ±0.04 mV or ±10%, whichever is greater
Other range: unspecified
Arrhythmia analysis: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Pacemaker: yes
RESP
Method: R-F(RA-LL) Impedance
Respiration rate:
Measurement and alarm range: 0~150 rpm
Resolution: 1 rpm
Measurement accuracy: ±2 rpm
Alarm accuracy: ±3 rpm
Apnea alarm: 10~40s
Scan speed: 12.5 mm/s, 25 mm/s
NIBP
Method: Oscillometry
Mode: Manual/AUTO/Continuous
Measurement interval in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480 minutes
Measurement period in Continuous mode: 5 minutes
Measurement and alarm range: 10 ~ 270 mmHg
Resolution: 1 mmHg
Cuff pressure accuracy: ±3 mmHg
Measurement accuracy:
Maximal mean deviation: ±5 mmHg
Maximal standard deviation: 8 mmHg
Over-pressure protection:
Adult mode: 297±3 mmHg
Pediatric mode: 240±3 mmHg
Neonatal mode: 147±3 mmHg
SpO2
Measurement and alarm range: 0 ~ 100%
Resolution: 1%
Measurement accuracy: 70%~100%: ±2%;
0%~69%: unspecified
PR
Measurement and alarm range: 30 ~ 250 bpm
Measurement accuracy: ±2 bpm or ±2%, whichever is greater
TEMP
Channel: single-channel
Measurement and alarm range: 0 ~ 50℃
Resolution: 0.1 ℃
Accuracy: ±0.1 ℃
Power supply
Adapter: 9 (V) DC
AC: 100 ~ 240 (V), 50/60 (Hz)
Maximal input power: 150VA
Safety classification: Class I, type CF defibrillation-proof applied part
Accessories
Adult fingertip SpO2 probe
SpO2 probe extension cable
Adult NIBP cuff
NIBP extension tube
ECG lead cable
ECG electrode
Temperature probe
Power cord
Power adapter
User Manual
Physical characteristic
Dimension: 192 mm(L) × 162 mm(W) × 243 mm(H)
Weight: 2.0 Kg