Test parameters: WBC 3-part differential counting, 24 parameters (including WBC, RBC, PLT color histogram);
Measurement principle: counting by electrical impedance method, colorimetric method to measure HGB;
Quality control method: L-J, automatic drawing and printing of quality control charts;
Sample volume: trace detection, 10μL peripheral blood or anticoagulant blood, pre-dilution mode 20μL;
Precision: WBC ≤ 3.0%, RBC ≤ 2.0%, HGB ≤ 2.0%, PLT ≤ 5.0%, MCV ≤ 0.5% ;
Accuracy: WBC ≤ ±6.0%, RBC ≤ ±5.0%, HGB ≤ ±4.0%, PLT ≤ ±10.0%, HCT ≤ ±5.0%;
Blank count: WBC ≤ 0.2×109/L, RBC ≤ 0.02×1012/L, HGB ≤ 1 g/L, PLT ≤ 10×109/L, HCT ≤±0.5%;
Carry-over: WBC ≤ 2.0%, RBC ≤ 1.0%, HGB ≤ 2.0%, PLT ≤ 2.0%;
Linearity deviation: WBC ≤ ±5%, RBC ≤ ±5%, HGB ≤ ±3%, PLT ≤ ±10%;
Linearity related coefficient: WBC ≥ 0.990, RBC ≥ 0.990, HGB ≥ 0.990, PLT ≥ 0.990;
Display: 8″ color LCD touch screen;
Detection channel: dual channel;
Test speed: 60 specimens/hour, continuous working for 24 hours;
Data storage: It can automatically store more than 300,000 groups of complete results (each result has three histograms);
Interface: USB interface,RS232 interface;
Power supply: ~100V-240V; 50/60Hz.