Introduction
Capnograph adopts mainstream method to measure the end-tidal carbon dioxide, airway respiration rate and other parameters. It can detect pulmonary ventilation function and reflect the circulation and pulmonary blood flow condition, as well as indirectly reflect the alveolar ventilation, which is applicable for use in pre-hospital emergency transfer, emergency department, ICU, operating room, neonatology, respiratory department and other occasions. The application population is adult, pediatric and neonate patients. The Capnograph connects with patient’ respiratory tract by the airway adapter and breathing tube.
Features
1)End-tidal CO2 measurement function
2)Respiration rate measurement function
3)Over-limit alarm function for End-tidal CO2 and respiration rate
4)Low battery alarm function
Performance
End-tidal CO2 measurement:
1)Measurement range: 0 ~150mmHg
2)Resolution: 1mmHg
3)Accuracy:
0~40 mmHg:±2 mmHg
41~70 mmHg:±5% of readings
71~100 mmHg:±8% of readings
101~150 mmHg:±10% of readings
Respiration rate measurement:
1)Measurement range: 2rpm~150rbpm
2)Resolution: 1rpm
3)Accuracy: ±1rpm
Safety characteristic:
1)Device type: internally powered equipment
2)Degree of protection against electric shock: type BF applied part
3)Degree of protection against ingress of liquid: IPX2
Accessories
User Manual
Adult gas path adapter
Neonate gas path adapter (optional)
Physical Characteristic
Dimension: 55×52×59mm
Weight: 97g (with batteries)
Working environment
Temperature: 5℃~40℃
Relative humidity: 30%~ 75%, no condensation
Atmospheric pressure: 700hPa~1060hPa
Power supply: DC 3V (2 AAA dry batteries)
Input power: ≤1VA